Comparación de la fiabilidad de "0.5" y "APEX" Marcado
de las medidas basadas en dos frecuencias, localizadores
apicales electrónicos
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
• Los localizadores electrónicos de ápice (Eals)
pueden determinar la longitud de trabajo más
conveniente y precisa.
•
El principio básico de estos Eals es que la conductividad
eléctrica del tejido que rodea el ápice de la raíz la cual es mayor
que en el interior del sistema de conductos radiculares.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
Se han fabricado muchos Eals a través de este principio, los fabricantes
afirman que sus productos pueden localizar la constricción apical.
97.5%
Root ZX (J. Morita Mfg Corp, Kyoto, Japón)
es un dispositivo que mide la impedancia de 2
frecuencias (0,4 y 8 kHz) al mismo tiempo y
calcula el cociente de las impedancias.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
0,4 KHz
8 kHz
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
De acuerdo con las instrucciones
del fabricante el localizador indica
"0,5" cuando la punta de la
lima se encuentra en la CA.
"0,5" longitud de trabajo ideal
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
Sin embargo, no todos los órganos dentarios tienen CA, incluso
pueden no tener,
principalmente con presencia de patología apical y reabsorción radicular.
En estos casos, sin CA, existe poca información en cuanto a qué
posición indica realmente la marca "0,5" en el localizador.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
Además, la marca "0,5" podría ser la CA o un área más allá de la CA según
otros estudios con Eals diferentes.
Gulabivala y Stock sugieren que Eals deben utilizarse para lograr "cero" para
obtener la mayor exactitud debido a las características de impedancia dado
que la CA no puede ser calculada con precisión.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
INTRODUCCIÓN
Los Eals deben obtener una medida fiable y exacta.
En estadística, una medida fiable significa medir algo
consistente con una variación mínima.
Sin embargo, la mayoría de los autores sólo se evalúa la
exactitud de Eals para determinar la longitud y no evaluar la
fiabilidad del método de medición.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
El presente estudio comparó la fiabilidad
de las mediciones "0.5" y "APEX"
obtenidas por 2 Eals.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
104
premolares inferiores
humanos extraídos se utilizaron.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
LAMINA DE ACRILICO
3 cm
3 cm
3 cm
6 cm
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
52 Dientes
52 Dientes
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
L1
LONGITUD 1
Lima Tipo K
#15 - 25
0.5
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
Calibrador Digital Mitutoyo
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
L2
LONGITUD 2
0.0
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
La L2 se restó de la longitud medida
previamente (L1),
el valor calculado (L3) se considera que la
distancia entre la "0,5" y marcas "APEX".
L2
L1
L3
LONGITUD 1
LONGITUD 1
Distancia
entre 0.5 y
Apex
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
El valor es positivo + se la lima se encuentra antes del foramen y negativo – si esta pasando el foramen
0.5
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
Fuji II LC en polvo-líquido es Gold Label 2 LC
Ionómero de Vidrio Restaurativo fotocurable
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
Discos SOF-LEX (3M ESPE)
3 a 4 mm
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
21 x
OPMI Pico microscope (Carl Zeiss)
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
MATERIALES Y METODOS
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
RESULTADOS
• En promedio, las medidas tomadas en el "0.5" marca por Root ZX y el i-Root
fueron consistentemente 0,26 y 0,29 mm antes del FM.
• Las dos marcas (0.5 y APEX) de los 2 Eals se puede utilizar
con fines clínicos
• No hubo diferencia significativa entre los Eals
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
DISCUSIÓN
La elección entre las 2 marcas ("0.5" y
"APEX") para la determinación de la
longitud de trabajo parece sobre todo
depender de la experiencia y la
preferencia del especialista.
En estudios microscópicos anteriores se ha informado que las
distancias entre la CA y foramen es de 0,5 mm.
Sin embargo, en este estudio, las distancias medidas entre "0.5" y APEX
fueron de solo 0,26 mm (Root ZX) y 0,29 mm i-Root) lo cual confirma que
los localizadores electrónicos no determinan exactamente .5mm antes
del foramen.
[email protected]
Jung I Y, Yoon B H, Lee S J. Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ Mark Measurements in Two Frequency-based Electronic Apex Locators. J Endod 2011;37:49-52.
Comparison of the Reliability of ‘‘0.5’’ and ‘‘APEX’’ in Electronic Apex Locators
DISCUSIÓN
En conclusión:
1. La marca "APEX" de los Eals utilizados en este estudio indico el Foramen
constantemente.
2. La marca "0,5" por Root ZX es de 0,26mm antes del foramen y con el i-Root es
de 0,29 mm antes del foramen.
3. Las dos marcas (0.5 y APEX) de los 2 Eals se puede utilizar con fines clínicos.